Title of article :
Comparison and validation of point of care lactate meters as a replacement for fetal pH measurement
Author/Authors :
Robert V. Ridenour، نويسنده , , Ravi P. Gada، نويسنده , , Brian C. Brost، نويسنده , , Brad S. Karon، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2008
Pages :
5
From page :
1461
To page :
1465
Abstract :
Objective To validate a point of care lactate device to replace fetal pH measurement. Study design and methods Cord blood samples drawn immediately following delivery were tested on the Nova Lactate Plus and ARKRAY Lactate Pro, the Corometrics 220 pH System, and the Vitros chemistry analyzer (used as lactate reference). Results Nova demonstrated a constant positive bias relative to the lactate reference method; while the Lactate Pro correlated well with the reference method up to 6 mmol/L. Receiver operating characteristic (ROC) curve analysis showed optimal sensitivity and specificity for predicting pH < 7.20 at lactate values of 6.8 mmol/L for the Nova and 4.8 mmol/L for the Lactate Pro. Conclusion Using Lactate Pro the best cut-off for predicting pH ≤ 7.20 was 4.8 mM; which coincides with current clinical cut-offs. Thus any lactate device that correlates well with the laboratory reference method can be used with a clinical cut-off of 4.8 mmol/L.