Title of article :
Sagittal curvature of total knee replacements predicts in vivo kinematics
Author/Authors :
Oliver Kessler، نويسنده , , Lutz Dürselen، نويسنده , , Scott Banks، نويسنده , , Henrich Mannel، نويسنده , , Frédéric Marin، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2007
Pages :
7
From page :
52
To page :
58
Abstract :
Background It is known that in vivo kinematics after total knee replacement is influenced by the design of the implant. The goal of this study was to show that the sagittal curvature of two different knee prostheses differing in geometric design predicts their in vivo motion behavior. Methods Three-dimensional tibio-femoral displacements of two prosthesis designs (single radius vs. dual radius) were measured during knee extension under weight bearing conditions by in vivo video fluoroscopy. Finite helical axes were computed to represent the tibio-femoral motions. Angular deviation α and the spatial localization deviation δ were used to characterize the motions. Angular deviation is the angle between each incremental finite helical axis and the medio-lateral axis of the femoral component of the prosthesis. The spatial localization deviation is the distance between each finite helical axis and the center of the femoral component of the prosthesis. Statistical comparisons were performed using the median and the interquartile range of the angular deviation and the spatial localization deviation. Findings The single-radius design showed finite helical axes concentrated at a single axis near to the medio-lateral axis of the femoral component. The angular and spatial localization deviation of the dual radius design were larger compared to the single radius design, exhibiting finite helical axes varying between two axes. Interpretation Video fluoroscopy in combination with finite helical axis analysis proved to be suitable methods to evaluate the in vivo kinematical behavior of total knee arthroplasty, which can be useful for implant designers. Knowledge of in vivo kinematics can also provide surgeons with more background information about the total knee arthroplasty models they implant.
Keywords :
Total knee arthroplasty , Single radius , knee kinematics , Video fluoroscopy , Helical axis analysis
Journal title :
Clinical Biomechanics
Serial Year :
2007
Journal title :
Clinical Biomechanics
Record number :
486667
Link To Document :
بازگشت