• Title of article

    A prototype system for testing force platform dynamic performance

  • Author/Authors

    P. S. Fairburn، نويسنده , , R. Palmer، نويسنده , , J. Whybrow، نويسنده , , S. Fielden، نويسنده , , S. Jones، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2000
  • Pages
    9
  • From page
    25
  • To page
    33
  • Abstract
    Previous work on in-situ calibration of force platforms has focused primarily on behaviour under static loading conditions. The prototype described in this paper is a dynamic system, designed to approximate the loading profile generated during human gait. The design, construction and performance of an oscillating lead pendulum assembly is described; including three-dimensional visual performance assessment and rigorous multi-component force analysis. No significant difference was found between measured data and a theoretical profile for vertical force and angular displacement. A significant difference was found for the horizontal force component. The authors believe that further development of this prototype model will allow accurate routine dynamic calibration.
  • Keywords
    Gait , Dynamic performance testing , Force platform
  • Journal title
    Gait and Posture
  • Serial Year
    2000
  • Journal title
    Gait and Posture
  • Record number

    487838