• Title of article

    Evaluation of an instrumented walkway for measurement of the kinematic parameters of gait

  • Author/Authors

    Robert G. Cutlip، نويسنده , , Corrie Mancinelli، نويسنده , , Frances Huber، نويسنده , , James DiPasquale، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2000
  • Pages
    5
  • From page
    134
  • To page
    138
  • Abstract
    The purpose of this study was to compare kinematic gait parameters measured with an instrumented walkway system (GAITRite®) and a video-based system (peak performance motus 3.1®). Subjects walked across a GAITRite mat with embedded pressure sensors. Reflective markers were attached to subjects’ shoes and video capture was simultaneously performed during each trial. Video data were then digitized manually using peak software. Correlation coefficients for all parameters measured with both systems were high (≥0.94). Significant differences between systems were found with analysis of variance (ANOVA) for two parameters, step length and stride velocity (P=0.003, 0.0002). The results of this study indicate that the instrumented walkway gave comparable results for temporal parameters but further investigation is needed to evaluate the fidelity of its spatial performance.
  • Keywords
    Gait , Kinematics , Motion analysis
  • Journal title
    Gait and Posture
  • Serial Year
    2000
  • Journal title
    Gait and Posture
  • Record number

    487847