Title of article :
A novel surface ionization source for ion mobility spectrometer
Author/Authors :
Hui-yong، Guo نويسنده , , Xiu-li، He نويسنده , , Xiao-guang، Gao نويسنده , , Jian، Jia نويسنده , , Jian-ping، Li نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2007
Abstract :
A surface ionization (SI) source is designed and prepared for ion mobility spectrometer (IMS). The source acts not only as an emitter but also an ion injector which can inject ions periodically into the drift region of drift tube. Using the dual-role source, the dimension of the drift tube can be decreased and the circuit for high voltage can be simplified efficiently. The IMS with the SI source has a response range of ~4 orders of magnitude and a good reproducibility to tri-ethylamine. Compared with radioactive ionization (RI), the ultra-short time for ion injection and the zero level base line of ion mobility spectrum are characteristics of the surface ionization.
Keywords :
Ion mobility spectrometer , Tri-ethylamine , Surface ionization
Journal title :
Analytica Chimica Acta
Journal title :
Analytica Chimica Acta