Title of article :
Flow-injection spectrophotometric determination of cobalt(II) at low µg l-1 levels with 4benzylpiperidinedithiocarbamate
Author/Authors :
Andac، M. نويسنده , , Asan، A. نويسنده , , Isildak، I. نويسنده , , Cesur، H. نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2001
Pages :
-142
From page :
143
To page :
0
Abstract :
A rapid, sensitive and reproducible flow-injection (FI) analysis method is described for the spectrophotometric determination of cobalt(II) using 4-benzylpiperidinedithiocarbamate as a chromogenic reagent. The method involves the measurement of the absorbance of the colored cobalt(II)-(4-benzylpiperidinedithiocarbamate)2 complex at 640nm. In the system, a microcolumn (MC) containing strong cation-exchange resin was placed between injection manifold and spectrophotometer to increase interactions between cobalt(II) and reagent, and for preconcentration of cobalt(II). The linear range of the proposed FI method was 0.6-100 µgl-1, and the limit of detection (signal/noise=3) was 0.1µgl-1 for a 10µl injection of cobalt(II) ion solution at a sampling rate of 60h-1. The relative standard deviation (R.S.D.) for 0.6-100 µgl-1 cobalt(II) was <1.0% (n=5). The precision and accuracy of the method were checked by analysis of certified reference metals, and the method was applied to river, sea and hot spring water samples.
Keywords :
Beverages analysis. , Potentiometry , Benzoate sensitive electrode
Journal title :
Analytica Chimica Acta
Serial Year :
2001
Journal title :
Analytica Chimica Acta
Record number :
49329
Link To Document :
بازگشت