Title of article :
Selenium determination in urine with atomic fluorescence detection
Author/Authors :
Sabe، Rosa نويسنده , , Rubio، Roser نويسنده , , Garcia-Beltr?n، Lydia نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2001
Pages :
-214
From page :
215
To page :
0
Abstract :
The aim of this study was to determine the Se concentration in urine samples under routine clinical laboratory conditions. The sample was completely mineralised using a focused microwave oven with a mixture of nitric acid and sulphuric acid for 14min. Se was measured by using hydride generation-atomic fluorescence spectrometry (HG-AFS). Complete recovery was achieved for SeCys, SeMet and TMeSe species. The method was validated by analysing the standard reference material SRM 2670 urine (normal and elevated levels), and by regular participation in Interlaboratory Exercises. Detection limit and precision in terms of reproducibility and repeatability were established. A study of the stability of digested samples and pre-reduced solutions over time was performed to propose the optimised method for routine analysis.
Keywords :
Benzoate sensitive electrode , Potentiometry , Beverages analysis.
Journal title :
Analytica Chimica Acta
Serial Year :
2001
Journal title :
Analytica Chimica Acta
Record number :
49364
Link To Document :
بازگشت