Title of article :
Sub-millimolar determination of formalin by pulsed amperometric detection
Author/Authors :
Surareungchai، Werasak نويسنده , , Ngamchana، Sirimarn نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2004
Pages :
-194
From page :
195
To page :
0
Abstract :
Formalin, formaldehyde in the presence of methanol, was determined by pulsed amperometric detection (PAD). A triple waveform using Edet=-0.3 V (tdet=30 ms), Eoxd=+0.8 V (toxd=200 ms), and Ered=-0.8 V (tred=350 ms) versus Ag/AgCl was applied at a Au electrode for detection in a flow injection (FI) system. The approach was rapid and yielded a sub-millimolar detection limit (0.0129 mM) with a dynamic range up to 100 mM. A precision of 8.8% R.S.D. at 1.0 mM for two hundred repetitive injections by the FI-PAD was obtained, whereas holding at a constant potential (-0.3 V versus Ag/AgCl) for anodic oxidation of formaldehyde caused the response to decrease dramatically after a few measurements. The method developed was used to analyze the formalin contents of water from rinsed samples of vegetables and fruit and ice-melt from seafood, and the method showed good agreement with the liquid chromatography (LC) method.
Keywords :
Pulsed amperometric detection , Flow injection analysis , formaldehyde , Formalin
Journal title :
Analytica Chimica Acta
Serial Year :
2004
Journal title :
Analytica Chimica Acta
Record number :
50167
Link To Document :
بازگشت