Title of article :
Kinetic spectrophotometric method for simultaneous determination of selenium and tellurium using partial least squares calibration
Author/Authors :
SHAMSIPUR، M. نويسنده , , Ghasemi، J. نويسنده , , Khajehsharifi، H. نويسنده , , Mousavi، M. F. نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2004
Pages :
-368
From page :
369
To page :
0
Abstract :
A kinetic-spectrophotometric method for the simultaneous determination of Se(IV) and Te(IV) is described. The method is based upon the catalytic effect of these cations on the reaction of toluidine blue with sulfide. Partial least squares calibration method was employed for the data manipulation and analysis. The concentrations were varied between 0.02–0.24 and 0.01–0.08 (mu)g/ml for Se(IV) and Te(IV), respectively. Cross-validation method was used to select the optimum number of factors (NF). The root mean square errors of difference for selenium and tellurium were 1.2 and 1.7 (mu)g/ml, respectively. Application of the method to artificial samples and several mixtures of standard solutions of Se(IV) and Te(IV) were performed and satisfactory results were obtained.
Keywords :
Simultaneous determination , selenium , Toluidine blue—sulfide catalysis , Spectrophotometry , PLS , Tellurium
Journal title :
Analytica Chimica Acta
Serial Year :
2004
Journal title :
Analytica Chimica Acta
Record number :
50261
Link To Document :
بازگشت