Title of article
Corrosion susceptibility of titanium covered by dental cements
Author/Authors
Y.L Turpin، نويسنده , , R.D Tardivel، نويسنده , , A Tallec، نويسنده , , A.C Le Menn، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2000
Pages
5
From page
57
To page
61
Abstract
Objectives: The aim of this study is to determine the influence of acidic fluorinated cements (as glass ionomer (GI) cements) on the passivation of titanium, using electrochemical investigations.
Methods: We realized experimental electrodes that associate titanium and dental cements. Polarization resistance of titanium electrodes has been determined for uncovered metal and electrodes covered with three dental cements. Studentʹs t-tests proved the reproducibility. We also compared successive voltammograms for uncovered titanium and GI covered titanium.
Results: Correct passivation was observed with zinc eugenate, but fluorinated GI or zinc phosphate coverage increased the corrosion susceptibility.
Significance: However there is no evidence of titanium depassivation when covered with cement. No clinical contraindication can be held.
Keywords
Dental cements , Polarization resistance , Titanium electrode