Title of article
Measurement of opalescence of tooth enamel
Author/Authors
Yong-Keun Lee، نويسنده , , Bin Yu، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2007
Pages
5
From page
690
To page
694
Abstract
Objectives
Opalescent dental esthetic restoratives look natural and esthetic in any light, react to light in the same manner as the natural tooth and show improved masking effect. The objective of this study was to determine the opalescence of tooth enamel with reflection spectrophotometers.
Methods
Color of intact bovine and human enamel was measured in the reflectance and transmittance modes. Two kinds of spectrophotometers were used for bovine and one kind was used for human enamel. The opalescence parameter (OP) was calculated as the difference in yellow-blue color coordinate (CIE Δb*) and red-green color coordinate (CIE Δa*) between the reflected and transmitted colors.
Results
Mean OP value of bovine enamel was 10.6 (±1.4) to 19.0 (±2.1), and varied by the configuration of spectrophotometers. Mean OP value of human enamel was 22.9 (±1.9).
Conclusions
Opalescence varied by the configuration of measuring spectrophotometer and the species of enamel. These values could be used in the development of esthetic restorative materials.
Keywords
OpalescenceEnamelReflected colorTransmitted color
Journal title
Journal of Dentistry
Serial Year
2007
Journal title
Journal of Dentistry
Record number
507672
Link To Document