Title of article
Coronary artery stenoses: A phantom study using contrast enhanced three-dimensional electron beam tomography
Author/Authors
Bin Lu، نويسنده , , RuPing Dai، نويسنده , , Yu-Hua Bai، نويسنده , , Sha He، نويسنده , , BaoLian Jing، نويسنده , , ShiLiang Jiang، نويسنده , , Nan Zhuang، نويسنده , , XiaoGang Sun، نويسنده , , Matthew J. Budoff، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2001
Pages
6
From page
95
To page
100
Abstract
This paper evaluated the accuracy of electron beam tomographic angiography (EBA) with conventional coronary arteriography (CCA) using four graded artificial stenoses in a postmortem swine coronary phantom model. The sensitivity, specificity, and accuracy of EBA for diagnosing significant stenosis (≥50% stenosis) were 94.3%, 96.7%, and 95.8%, respectively. The diagnostic accuracy of EBA had no significant difference with CCA (χ2=0.0162; P>.05). EBA three-dimensional (3D) procedures had high interobserver reproducibility (k=.92–.95, P>.05). Maximum intensity projection (MIP) was the most sensitive and curved planar reformation (CPR) was the most accurate 3D procedure for quantitatively identifying coronary stenosis. EBA yields promising results concerning the visualization of coronary artery stenosis with high accuracy for stenoses >50%.
Keywords
Electron-beam , Coronary Artery Disease , angiography , three-dimensional , Phantom study , computed tomography
Journal title