    Improved Detection Limits and Sensitivities of Potentiometric Titrations

  • Author/Authors

    Pretsch، erno نويسنده , , Bakker، Eric نويسنده , , Peper، Shane نويسنده , , Ceresa، Alan نويسنده ,

  • Issue Information
    دوهفته نامه با شماره پیاپی سال 2001
  • Pages
    -3767
  • From page
    3768
  • To page
    0
  • Abstract
    Zero-current ion fluxes through polymer membranes of ion-selective electrodes (ISEs) may lead to biased endpoints of potentiometric titrations. The bias is eliminated, and the sensitivity of the end-point detection is improved through reducing transmembrane ion fluxes by an appropriate choice of the inner solution. Surprisingly, ISE membranes that have a significant primary ion flux toward the inner solution show much larger sensitivities than expected by titration theory; however, depending on the experimental conditions, their application may bias the endpoint. With the optimal systems, endpoint detection is now possible with total sample concentrations below 10^-6 M, as demonstrated by the titration of EDTA with a Pb^2+ solution.