Title of article :
Infrared Multiphoton Dissociation with a Hollow Fiber Waveguide
Author/Authors :
Hannis، James C. نويسنده , , Drader، Jared J. نويسنده , , Hofstadler، Steven A. نويسنده ,
Issue Information :
دوهفته نامه با شماره پیاپی سال 2003
Abstract :
A novel scheme for performing infrared multiphoton dissociation (IRMPD) is presented in which a hollow fiber waveguide (HFWG) is used to transmit IR radiation into the ion storage region of a mass spectrometer. Efficient dissociation of oligonucleotide and protein ions is demonstrated on an ESI-FTICR instrument in which IRMPD is performed in the external ion reservoir and on a quadrupole ion trap. Using a simple optical scheme consisting of a single focusing lens and an x, y translator, the 10.6-(mu)m IR laser beam, initially 3.5 mm in diameter, is focused into the vacuum-sealed HFWG. The small internal diameter and the high transfer efficiency of the waveguide allow IR radiation of high power density to be employed for IRMPD. In studies performed on a quadrupole ion trap, a 500-(mu)m-i.d. waveguide was used as a medium to transmit IR radiation directly through a 700-(mu)m orifice in the ring electrode. Efficient IRMPD of both a 12-mer oligonucleotide and the protein melittin were performed at laser powers of 0.5 and 3.2 W, respectively.
Keywords :
gas_phase measurement , particle_phase measurement
Journal title :
Analytical Chemistry
Journal title :
Analytical Chemistry