Title of article :
Self-Referencing Ceramic-Based Multisite Microelectrodes for the Detection and Elimination of Interferences from the Measurement of L-Glutamate and Other Analytes
Author/Authors :
Gerhardt، Greg A. نويسنده , , Burmeister، Jason J. نويسنده ,
Issue Information :
دوهفته نامه با شماره پیاپی سال 2001
Abstract :
A self-referencing technique utilizing two microelectrodes on a ceramic-based multisite array is employed for confirmation and elimination of interferences detected by enzyme-based microelectrodes. The measurement of L-glutamate using glutamate oxidase was the test system; however, other oxidase enzymes such as glucose oxidase can be employed. One recording site was coated with Nafion with L-glutamate oxidase and bovine serum albumin (BSA) cross-linked with glutaraldehyde while the other had Nafion with BSA cross-linked with glutaraldehyde. Differences in the chemistry of the two recording sites allowed for identification and elimination of interfering signals to be removed from the analyte response. The electrode showed low detection limits (LOD = 0.98 ± 0.09 muM, signal-to-noise ratio of 3), fast response times (T90 ~1 s), and excellent linearity (R2 = 0.999 ± 0
Keywords :
Isoflavone glycoside 6"-O-acetates , High-performance liquid chromatography-mass spectrometry (LC-MS) , Solid-phase extraction (SPE) , Trifolium pratense , Isoflavone glycoside 6"-O-malonates , red clover
Journal title :
Analytical Chemistry
Journal title :