Title of article
Picomolar Detection Limits with Current-Polarized Pb2+ Ion-Selective Membranes
Author/Authors
Pergel، Eva نويسنده , , Gyurcs?nyi، Robert E. نويسنده , , T?th، Kl?ra نويسنده , , Lindner، Ern? نويسنده ,
Issue Information
دوهفته نامه با شماره پیاپی سال 2001
Pages
-4248
From page
4249
To page
0
Abstract
Minor ion fluxes across ion-selective membranes bias submicromolar activity measurements with conventional ion-selective electrodes. When ion fluxes are balanced, the lower limit of detection is expected to be dramatically improved. As proof of principle, the flux of lead ions across an ETH 5435 ionophore-based lead-selective membrane was gradually compensated by applying a few nanoamperes of galvanostatic current. When the opposite ion fluxes were matched, and the undesirable leaching of primary ions was eliminated, Nernstian response down to 3 × 10-12 M was achieved.
Keywords
Solid-phase extraction (SPE) , Trifolium pratense , Isoflavone glycoside 6"-O-malonates , red clover , Isoflavone glycoside 6"-O-acetates , High-performance liquid chromatography-mass spectrometry (LC-MS)
Journal title
Analytical Chemistry
Serial Year
2001
Journal title
Analytical Chemistry
Record number
51862
Link To Document