Title of article :
Electroencephalographic recording during transcranial magnetic stimulation in humans and animals
Author/Authors :
John R. Ives، نويسنده , , Alexander Rotenberg، نويسنده , , Roberto Poma، نويسنده , , Gregor Thut، نويسنده , , Alvaro Pascual-Leone، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2006
Pages :
6
From page :
1870
To page :
1875
Abstract :
Objective We report on the development of an EEG recording system, comprised of electrodes and amplifiers that are compatible with TMS (single and rapid-rate) in both human and animal studies. Methods We assembled a versatile multi-channel EEG recording system consisting of: (1) two types of electrodes that are safe during TMS or rTMS. (2) Low slew-rate EEG amplifiers that recover within a few milliseconds after the application of TMS pulses. Results The two electrode types: (a) a conductive-plastic surface electrode with a conductive-silver epoxy coat and (b) a subdermal silver wire electrode (SWE) are compatible to TMS pulses. The amplifiers recover within 30 ms, so that the EEG can be viewed online, essentially without interruption and/or blocking or excessive artifact. Conclusions Our TMS compatible electrode and EEG recording system allows safe and online viewing/recording of the subjectʹs (human or animal) EEG/EP during experiments or studies involving TMS or rTMS applications. The TMS compatible electrode/amplifier system can be used with any EEG recording instrument. Significance A simple recording technique coupled with new electrodes permit safe and readable EEG records during TMS in humans and animals. Such online monitoring of the EEG would allow control of TMS/rTMS parameters based on EEG activity.
Keywords :
human , safety , Transcranial magnetic stimulation (TMS) , Animal , electrode , Electroencephalogram (EEG) , Epilepsy
Journal title :
Clinical Neurophysiology
Serial Year :
2006
Journal title :
Clinical Neurophysiology
Record number :
523657
Link To Document :
بازگشت