Title of article
Implantation of transvenous nonthoracotomy cardioverter-defibrillator systems in patients with permanent endocardial pacemakers
Author/Authors
Ross Brooks، نويسنده , , Hasan Garan، نويسنده , , Brian A. McGovern، نويسنده , , Jeremy N. Ruskin، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1995
Pages
9
From page
45
To page
53
Abstract
Among 177 patients in whom a nonthoracotomy approach was initially used to implant a cardioverter-defibrillator system, 11 (6%) patients also received a separately implanted permanent pacemaker. The main problem encountered in these patients were previously implanted unipolar pacemakers (n = 3) and ventricular pacing leads positioned at the right ventricular apex, the latter interfering with optimal placement of the tripolar implantable cardioverter-defibrillator (ICD) lead (n = 9). The approaches used to solve these problems were individualized and included placement of the ICD sensing lead at the right ventricular outflow tract (n = 3), initial placement (n = 1) or subsequent repositioning (n = 2) of the right ventricular pacing lead at the outflow tract, upgrade from unipolar to bipolar systems (n = 2), reprogramming from the DDD to AAI mode (n = 2), inactivation of the pacemaker (n = 1), and simultaneous placement of a single-chamber atrial pacemaker with the ICD lead (n = 2). These revisions fulfilled the pacing needs in each patient and prevented unfavorable sensing interaction between the two systems.
Journal title
American Heart Journal
Serial Year
1995
Journal title
American Heart Journal
Record number
526375
Link To Document