Title of article :
Implantation of transvenous nonthoracotomy cardioverter-defibrillator systems in patients with permanent endocardial pacemakers
Author/Authors :
Ross Brooks، نويسنده , , Hasan Garan، نويسنده , , Brian A. McGovern، نويسنده , , Jeremy N. Ruskin، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1995
Pages :
9
From page :
45
To page :
53
Abstract :
Among 177 patients in whom a nonthoracotomy approach was initially used to implant a cardioverter-defibrillator system, 11 (6%) patients also received a separately implanted permanent pacemaker. The main problem encountered in these patients were previously implanted unipolar pacemakers (n = 3) and ventricular pacing leads positioned at the right ventricular apex, the latter interfering with optimal placement of the tripolar implantable cardioverter-defibrillator (ICD) lead (n = 9). The approaches used to solve these problems were individualized and included placement of the ICD sensing lead at the right ventricular outflow tract (n = 3), initial placement (n = 1) or subsequent repositioning (n = 2) of the right ventricular pacing lead at the outflow tract, upgrade from unipolar to bipolar systems (n = 2), reprogramming from the DDD to AAI mode (n = 2), inactivation of the pacemaker (n = 1), and simultaneous placement of a single-chamber atrial pacemaker with the ICD lead (n = 2). These revisions fulfilled the pacing needs in each patient and prevented unfavorable sensing interaction between the two systems.