• Title of article

    Early and late assessment of the Micro Stent PL coronary stent for restenosis and suboptimal balloon angioplasty, ,

  • Author/Authors

    John G. Webb، نويسنده , , Jeffrey J. Popma، نويسنده , , Alexander J. Lansky، نويسنده , , Ronald G. Carere، نويسنده , , Alan Rabinowitz، نويسنده , , Joel Singer، نويسنده , , Arthur Dodek، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1997
  • Pages
    6
  • From page
    369
  • To page
    374
  • Abstract
    This prospective study represents the initial assessment of the Micro Stent PL (Arterial Vascular Engineering, Inc.) coronary stent. From one to three radiopaque stainless steel stents, each measuring 4 mm long, were premounted onto specially designed balloon catheters. A total of 123 stents were implanted in 41 patients without procedural failure or complications. Stent dislodgment proved a concern, with 7 of 123 stents (5.7%) moving >3 mm from the site of placement and late downstream migration occurring in an additional patient. Subacute stent thrombosis occurred in two patients (5%). Six-month angiographic follow-up was available in 37 of 41 patients (90%). Minimal lumen diameter at baseline was 0.93 ± 0.51 mm, increasing to 2.74 ± 0.49 mm after stenting, and falling to 1.66 ± 0.89 mm at 6 months; this represents a late loss of 60% of the initial gain. Restenosis, based on a binary definition of >50% diameter stenosis, was documented in 18 patients (49%). Advantages of the Micro Stent PL include its radiopacity and marked ease of distal delivery. The potential for stent dislodgment has implications for future stent designs. The role of the Micro Stent PL in managing restenosis is unclear, but it appears useful in the management of dissection and threatened closure after balloon angioplasty. (Am Heart J 1997;133:369-74.)
  • Journal title
    American Heart Journal
  • Serial Year
    1997
  • Journal title
    American Heart Journal
  • Record number

    530849