Title of article :
Mechanical recanalization of total coronary occlusions with the use of a new guide wire, ,
Author/Authors :
Bernhard Reimers، نويسنده , , Nino Camassa، نويسنده , , Carlo Di Mario، نويسنده , , Tatsuro Akiyama، نويسنده , , Lucia Di Francesco، نويسنده , , Leo Finci، نويسنده , , Antonio Colombo، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1998
Pages :
6
From page :
726
To page :
731
Abstract :
The mechanical approach in the recanalization of total coronary occlusions consisted of the use of a new 0.014-inch standard coronary guide wire with jointless spring coil design that improves steering characteristics and tip stiffness. In addition, a 0.014-inch soft tip wire with hydrophilic coating and low-profile 1.5 mm over-the-wire balloons were used. The first wire was used selectively in 86 patients to treat 95 total occlusions, of which 51 (54%) were older than 3 months. Unfavorable angiographic characteristics were present in 79 (83%) of 95 lesions. Overall crossing success was 71% (67 of 95 lesions). Complications were one coronary perforation with cardiac tamponade necessitating emergency bypass surgery. In conclusion, the mechanical approach with the use of the standard coronary guide wire with jointless spring coil design provides a high success rate in the recanalization of unfavorable total occlusions. (Am Heart J 1998;135:726-31.)
Journal title :
American Heart Journal
Serial Year :
1998
Journal title :
American Heart Journal
Record number :
531177
Link To Document :
بازگشت