Title of article
Incidence, predictors, and outcomes of device failure of X-sizer thrombectomy: Real-world experience of 200 cases in 5 years
Author/Authors
Chi Hang Lee، نويسنده , , Huay Cheem Tan، نويسنده , , Hwee-Bee Wong، نويسنده , , Xiao-Ling Zhang، نويسنده , , Shermaine Fun، نويسنده , , Mae Gay، نويسنده , , Lucy Qu، نويسنده , , Jimmy Lim، نويسنده , , Adrian Low، نويسنده , , Yean Teng Lim، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2007
Pages
1
From page
14
To page
14
Abstract
Background
Clinical studies have evaluated the safety and efficacy of adjunctive X-sizer (EndiCOR Inc, San Clemente, CA) in percutaneous coronary intervention (PCI). However, patient and lesion subsets were highly selected, and extrapolation of the results to daily practice is problematic.
Methods
X-sizer thrombectomy was performed in 200 procedures from August 2000 to July 2005. The indications for the procedures were primary PCI in 71%. Device failure was defined as the occurrence of 1 or more of the following conditions: (a) failure of the X-sizer to reach the target segment, (b) failure to achieve final thrombolysis in myocardial infarction 3 flow, (c) slow flow or no-reflow, (d) distal embolization, and (e) coronary perforation.
Results
Device failure occurred in 48 procedures, giving a device failure rate of 24%. Logistic regression analysis showed that ostial lesion was the only independent predictor of device failure (OR 4.89, 95% CI 1.23-19.51, P = .024). A total of 19 patients had developed 22 adverse events, giving a 30-day adverse event rate of 9.5%. Among these, there were 13 deaths, 4 strokes, 2 reinfarctions, 2 repeat PCIs, and 1 coronary artery bypass grafting. Logistic regression analysis showed that X-sizer device failure was independently associated with 30-day adverse events (OR 3.42, 95% CI 1.04-11.25, P = .043).
Conclusions
The incidence of device failure of X-sizer thrombectomy was 24%. Ostial lesion was an independent predictor of device failure. Notably, X-sizer device failure was independently associated with 30-day adverse event. These highlight the importance of case selection and the problem with its use in ostial lesions.
Journal title
American Heart Journal
Serial Year
2007
Journal title
American Heart Journal
Record number
534717
Link To Document