• Title of article

    Automated external defibrillation versus manual defibrillation for prolonged ventricular fibrillation: Lethal delays of chest compressions before and after countershocks

  • Author/Authors

    Robert A. Berg، نويسنده , , Ronald W. Hilwig، نويسنده , , Karl B. Kern، نويسنده , , Arthur B. Sanders، نويسنده , , Lyndon C. Xavier، نويسنده , , Gordon A. Ewy، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2003
  • Pages
    10
  • From page
    458
  • To page
    467
  • Abstract
    Study objective We sought to determine whether the delays in chest compressions and defibrillation associated with an automated external defibrillator would adversely affect outcome compared with manual defibrillation in a swine model of out-of-hospital prolonged ventricular fibrillation. Methods After 8 minutes of untreated ventricular fibrillation, 16 swine (33±4 kg) were randomly assigned to automated external defibrillator defibrillation or manual defibrillation with the same biphasic truncated exponential waveform 150-J shock through the same type of pads. Defibrillation with the automated external defibrillator was performed as recommended by the manufacturer, and manual defibrillation was provided per American Heart Association Guidelines. The primary outcome measure was 24-hour survival with good neurologic outcome. Data are described as means±SD. Results None of 8 animals in the automated external defibrillator group survived for 24 hours, whereas 5 of 8 animals in the manual defibrillation group survived 24 hours, all with good neurologic outcome (P=.027). The time interval from simulated defibrillator arrival to first compressions was 98±18 seconds in the automated external defibrillator group versus 68±15 seconds in the manual defibrillation group. In particular, the interval from first shock to first chest compressions was 46±18 seconds versus 22±16 seconds, respectively. The mean percentage of time that chest compressions were performed in the first minute after the first countershock was 15%±13% versus 40%±15%, respectively. As a result, return of spontaneous circulation within 5 minutes of simulated defibrillator arrival occurred in only 1 of 8 animals in the automated external defibrillator group versus 6 of 8 animals in the manual defibrillation group. Conclusion The longer delays in chest compressions with automated external defibrillator defibrillation versus manual defibrillation can worsen the outcome from prolonged ventricular fibrillation
  • Journal title
    Annals of Emergency Medicine
  • Serial Year
    2003
  • Journal title
    Annals of Emergency Medicine
  • Record number

    537422