• Title of article

    Acute endotracheal tube occlusion caused by use of an esophageal detector device: Report of a case and a discussion of its utility

  • Author/Authors

    Robert S Wright، نويسنده , , Theodore J Burdumy، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2004
  • Pages
    4
  • From page
    626
  • To page
    629
  • Abstract
    An experienced intensivist performed an emergency endotracheal intubation on a 72-year-old woman without difficulty. The vocal cords were visualized during the intubation. An esophageal detector device was used to confirm correct airway placement. After aspiration of the device, it was impossible to easily ventilate the patient. The endotracheal tube was removed and found to be occluded by a mucus plug. It is postulated that aspiration of the syringe on the esophageal detector device suctioned the mucus plug from the bronchial tree into the endotracheal tube and precluded ventilation of the airway. This complication has not been previously reported. The utility and possible hazards of the esophageal detector device are discussed.
  • Journal title
    Annals of Emergency Medicine
  • Serial Year
    2004
  • Journal title
    Annals of Emergency Medicine
  • Record number

    537644