• Title of article

    Dependence of photon interaction depth on linear attenuation coefficient in high pure germanium detectors

  • Author/Authors

    Haluk Yucel، نويسنده , , M. Atif Cetimagener، نويسنده , , Halimagel Demimagerel، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1996
  • Pages
    7
  • From page
    535
  • To page
    541
  • Abstract
    The present work is concerned with the construction of a semiempirical function de(μ, Eγ) of the effective photon interaction depth depending upon both the linear attenuation coefficient of Ge and γ-ray energy. The interaction depth results obtained from the measurements in two high purity germanium crystals, which are nearly equal in volume, have been fitted by using a function. The semiempirical function agrees with the measured interaction depth values on the average to within 3–4%. The knowledge of effective interaction depth of γ-rays in germanium detector is a useful parameter in order to calculate the absolute efficiency for any source-to-detector distance on the particular detector.
  • Journal title
    Applied Radiation and Isotopes
  • Serial Year
    1996
  • Journal title
    Applied Radiation and Isotopes
  • Record number

    539513