Title of article
Progress towards an alanine/ESR therapy level reference dosimetry service at NPL
Author/Authors
P. H. G. Sharpe، نويسنده , , K. Rajendran، نويسنده , , J. P. Sephton، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1996
Pages
5
From page
1171
To page
1175
Abstract
This paper describes work being carried out at the National Physical Laboratory towards the establishment of an alanine reference dosimetry service for radiotherapy applications. A precision fused quartz holder has been constructed to allow precise positioning of alanine dosimeters in the ESR cavity. A novel method of signal analysis based on spectrum fitting has been developed to minimize the effect of baseline distrotions. Data are also presented on the relative response of alanine to 60Co gamma rays and high energy photons (4–12 MeV).