Title of article :
Alanine-ESR In Vivo dosimetry: a feasibility study and possible applications
Author/Authors :
F. Kuntz، نويسنده , , J. Y. Pabst، نويسنده , , J. P. Delpech، نويسنده , , J. P. Wagner، نويسنده , , E. Marchioni، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1996
Pages :
6
From page :
1183
To page :
1188
Abstract :
A new alanine-ESR dosimeter has been developed at AERIAL in order to study its potential use in radiotherapy. Alanine-ESR results are compared with ion chamber for depth-dose measurements. A good concordance has been found between provisional dosimetry and absorbed dose during high dose rate and intra operative treatments. The results of the experiments indicate that alanine-ESR dosimetry is suited to check dose optimisation routines and seems to be a promising in vivo dosimetry technique.
Journal title :
Applied Radiation and Isotopes
Serial Year :
1996
Journal title :
Applied Radiation and Isotopes
Record number :
539619
Link To Document :
بازگشت