Title of article
A unique calibration curve for alanine EPR dosimetry systems
Author/Authors
A. Wieser، نويسنده , , R. Girzikowsky، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1996
Pages
7
From page
1269
To page
1275
Abstract
The alanine EPR dose-response curve was proposed to follow a saturating exponential function. In this study the characteristic parameter D373of the exponential function was estimated for two types of dosimeters and different EPR spectrometer configurations. The results suggest that only below about 0.3 mW the alanine EPR signal is not saturated with microwave power. A unique alanine calibration curve is suggested for measurements with non-saturating microwave power.
Journal title
Applied Radiation and Isotopes
Serial Year
1996
Journal title
Applied Radiation and Isotopes