Title of article
Validation of the limulus amebocyte lysate (LAL) test for routine PET radiopharmaceuticals
Author/Authors
S. Zijlstra، نويسنده , , P. Gerken، نويسنده , , C. Rechin، نويسنده , , R. Wortmann، نويسنده , , G. Notohamiprodjo، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1997
Pages
4
From page
51
To page
54
Abstract
The kinetic turbidimetric limulus amebocyte lysate test was validated as method for detecting endotoxins in short-lived radiopharmaceutical samples. Using this method, radiopharmaceuticals can be released for administration to humans after the test, without extensive loss of radioactivity. Inhibition or enhancement on the LAL results by the product samples were examined in more detail and eliminated.