• Title of article

    Measurements with Chinese LiF:Mg,Cu,P (GR 200) TL detectors using a volume-type oven

  • Author/Authors

    U. A. Fill، نويسنده , , D. F. Regulla، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1997
  • Pages
    2
  • From page
    133
  • To page
    134
  • Abstract
    Chinese LiF:Mg,Cu,P (GR-200) detectors with the advantages of high sensitivity and other favourable dosimetric properties require a restricted temperature treatment for read-out and annealing, including high heating-up and cooling-down rates. Modifications to a conventional readily-available commercial volume-type oven are described that achieve a precision of ≤% (1σrel) for doses of 50–1000 μGy, a range which is of significance for environmental, occupational and patient diagnostic dosimetry.
  • Journal title
    Applied Radiation and Isotopes
  • Serial Year
    1997
  • Journal title
    Applied Radiation and Isotopes
  • Record number

    539728