Title of article :
75Se gamma ray intensity measurements
Author/Authors :
L. C. Longoria، نويسنده , , J. S. Benitez، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1997
Pages :
6
From page :
1069
To page :
1074
Abstract :
A study was made of the 75Se gamma ray spectrum using a HPGe detector. The detector was calibrated with a set of standard sources and the intensities of 13 75Se gamma lines were measured. Correction factors for source holder absorption, coincidence and random summing were also applied. The estimated uncertainty was less than 1% for the more intense photons.
Journal title :
Applied Radiation and Isotopes
Serial Year :
1997
Journal title :
Applied Radiation and Isotopes
Record number :
539872
Link To Document :
بازگشت