• Title of article

    Evaluation of a microtomography system with an X-ray microfocus tube

  • Author/Authors

    R. T. Lopes، نويسنده , , J. L. Rodrigues، نويسنده , , J. T. De Assis، نويسنده , , E. F. O. De Jesus، نويسنده , , L. F. De Oliveira، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1997
  • Pages
    6
  • From page
    1437
  • To page
    1442
  • Abstract
    The microfocus radiography system has been realized by means of a microfocus X-ray tube (maximum 160 kV), an image intensifier data collector and a five degree of freedom manipulation table. An 8-bit image acquisition PC-card and software were included to compose the microtomography system. To assemble the system, an image quantitative analysis was carried out by evaluation of the focal spot size, modulation transfer function of the system, and the ‘defect discernibility curves’. The 2D microtomographies were carried out using the mean of 10 image central lines in each projection. Image processing techniques were also used to obtain better results. The image reconstruction program is based on a filtered backprojection algorithm using a special window. Images of various types of samples were carried out in order to verify the performance of the system.