Title of article
Neutron-gamma well logging apparatus with more sensitivity for chlorine
Author/Authors
A. I. Kedrov، نويسنده , , N. V. Popov، نويسنده , , A. S. Khimchenko، نويسنده , , A. S. Tsybin، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1997
Pages
2
From page
1647
To page
1648
Abstract
A new neutron-gamma well logging apparatus for separating in the earth section of collection strata with poor mineralized water is described. In such a setting construction materials other than iron are used. The probing device (sonde) uses a Cf252 neutron source and NaI(Tl) spectrometric detector. In conditions of a closed oil well in the optimum mineralization range (15–40 g chloride salt/l water) the difference in registrated data against poorly minearlized water with fresh fluid is more than 10% (in porosity 20%). The apparatus has an external strong frame with diameter of 90 mm. the maximum operating temperature is 120°C. The apparatus is designed for investigation of oil, gas, hydrogeological wells and ecological purposes.
Journal title
Applied Radiation and Isotopes
Serial Year
1997
Journal title
Applied Radiation and Isotopes
Record number
539956
Link To Document