• Title of article

    Body composition stability in Lunar DPX

  • Author/Authors

    Russell H. Nord، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1998
  • Pages
    2
  • From page
    519
  • To page
    520
  • Abstract
    This study examines how well a Lunar DPX can compensate for certain aging changes in its X-ray source. Total body scans were done on a phantom having variable soft tissue composition. Aging of the X-ray source was simulated by placing a thin aluminum foil over the beam exit window, and by deliberately misadjusting the high voltage power supply. The % fat measurement was found to change by up to 12%, but returned to within about 2% after calibration.
  • Journal title
    Applied Radiation and Isotopes
  • Serial Year
    1998
  • Journal title
    Applied Radiation and Isotopes
  • Record number