Title of article
Total body chlorine: Calibration of the in vivo neutron activation measurement
Author/Authors
J. R. Groenewegan and R. M.A. Loyns، نويسنده , , S. Yasumura، نويسنده , , R. I. Moore، نويسنده , , X. Zhao، نويسنده , , H. M. Rarback، نويسنده , , A. F. Lomonte، نويسنده , , K. A. Vodopia، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1998
Pages
3
From page
533
To page
535
Abstract
Total body chlorine (TBCl), used to estimate the extracellular space, is measured by delayed-gamma neutron activation (DGNA) using the reaction 37Cl(n,γ)38Cl, at Brookhaven National Laboratory. During the calibration process, we noticed that different values were obtained when different amounts of Cl were placed in the phantom. This non-linear relationship is due to the thermal neutron flux suppression by the thermal neutron capture reaction 35Cl(n,γ)36Cl. Monte Carlo simulations confirm the results of phantom measurements showing an inverse relationship between the Cl content in the phantom and the γ-ray yield per gram Cl. Thus, it is important to calibrate the DGNA system for TBCl using phantom standards containing an amount of Cl close to that expected in the individual undergoing measurement.
Journal title
Applied Radiation and Isotopes
Serial Year
1998
Journal title
Applied Radiation and Isotopes
Record number
540055
Link To Document