Title of article :
Calibration of a prototype in vivo total body composition analyser using 14 MeV neutron activation and the associated particle technique
Author/Authors :
S. Mitra، نويسنده , , J. E. Wolff، نويسنده , , R. Garrett، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1998
Pages :
3
From page :
537
To page :
539
Abstract :
A prototype in vivo total body composition analyser has been constructed for determining the total body contents of nitrogen (TBN), carbon (TBC) and oxygen (TBO) in young experimental animals such as sheep or pigs by 14 MeV neutron activation using a commercially available associated particle sealed tube neutron generator (APSTNG). The instrument was calibrated by scanning phantoms of different sizes in the mass range 10–36 kg, filled with a mixture of elements as found in a normal human body. Good agreement was found between the measured and expected values of N, O and C when two phantoms of similar dimensions but of different composition were scanned. With four 15 cm × 15 cm cross section and 45 cm long NaI(Tl) gamma detectors and a radiation dose of ≈ 20 μSv due to neutrons, the expected precisions for a 28 kg animal, CV% (based on counting statistics) are N: 9.3, C: 2.3 and O: 1.4.
Journal title :
Applied Radiation and Isotopes
Serial Year :
1998
Journal title :
Applied Radiation and Isotopes
Record number :
540056
Link To Document :
بازگشت