Title of article :
Bioelectrical impedance spectroscopy (BIS) in young children with acute and semi-acute hydration disorders: Potentials and limitations
Author/Authors :
Manolo Mazariegos، نويسنده , , Christiane Pithan، نويسنده , , Andrea Meyer، نويسنده , , Ivan Mendoza، نويسنده , , PETER FURST، نويسنده , , Noel W. Solomons، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1998
Pages :
4
From page :
611
To page :
614
Abstract :
This study aimed to evaluate the feasibility and reliability of bioelectrical impedance spectroscopy in young children suffering of acute hydrational disorders. Whole body and segmental measurements were carried out in a group of 42 of children aged 4 to 147 months, using a BIS analyzer (Xitron 4000B). This phase of the study involved several hundred of BIS measurement, which showed the feasibility of using this technique in young children. Using the sweep mode and the modeling software provided for the manufacturer of the instrument, the resistance of the extracellular (RECW) and intracellular (RICW) compartment was calculated. Correlation coefficient (r) and technical error of measurement (TEM) were obtained from paired BIS measurements. RECW showed to be highly reproducible (r = 0.99), with a relatively low TEM (from 1 to 3.5%), across all segments. The reliability was markedly lower respect to RICW, which shows the potential application of BIS technique even in critically ill young child populations.
Journal title :
Applied Radiation and Isotopes
Serial Year :
1998
Journal title :
Applied Radiation and Isotopes
Record number :
540077
Link To Document :
بازگشت