Title of article :
In vivo X-ray fluorescence (XRF) measurement of uranium in bone
Author/Authors :
J. M. OʹMeara، نويسنده , , D. R. Chettle، نويسنده , , F. E. McNeill، نويسنده , , C. E. Webber، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1998
Abstract :
This study investigates the applicability of X-ray fluorescence (XRF) to measuring bone uranium concentrations, using a 57Co source to excite the uranium X-rays, with the source and detector in an approximate 180° backscatter geometry relative to the sample position. It is demonstrated, by experiment and Monte Carlo simulation, that the X-ray to coherent peak ratio is linearly related to concentration and is independent of variations in source-sample geometry, thickness of overlying tissue and tibia size. Preliminary in vivo measurements indicate a minimum detectable concentration (MDC) of approximately 20 μg/g, which may not be sufficiently sensitive for monitoring occupational workers. However, a larger study of occupationally exposed individuals as well as work with subjects with known significant accidental uranium exposures is necessary to assess the clinical usefulness of this system.
Journal title :
Applied Radiation and Isotopes
Journal title :
Applied Radiation and Isotopes