Title of article :
Si(Li) detector response for low-energy photons
Author/Authors :
Marie-Christine Lépy، نويسنده , , Johann Plagnard، نويسنده , , Philippe Stemmler، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1998
Abstract :
Monochromatized synchrotron radiation was employed to calibrate a silicon–lithium detector in the 1–7 keV energy range. The efficiency calibration was achieved using a reference proportional counter. The absorbing layer thickness was determined by fitting theoretical efficiency functions to experimental data; the silicon “dead layer” appears to operate as a partially active layer. The peak shape calibration is established by fitting an Hypermet-type function to monochromatic photon spectra. Knowledge of the relevant shape parameters dependence on the energy allows improved processing of complex X-ray spectra.
Journal title :
Applied Radiation and Isotopes
Journal title :
Applied Radiation and Isotopes