Title of article :
Active measurements of indoor concentrations of radon and thoron gas using charcoal canisters
Author/Authors :
K. N. Yu، نويسنده , , Z. J. Guan، نويسنده , , E. C. M. Young، نويسنده , , M. J. Stokes، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1998
Abstract :
Standardized charcoal canisters with diameters of 4 inch, of a type recommended by the USEPA for passive measurement of radon gas concentrations, have been modified for active air sampling. Simultaneous measurement of the concentrations of radon (Rn) and thoron (Tn) in air can be obtained by counting the areas under γ-ray peaks. Sample results obtained in a laboratory and in a residence with different conditions are also given. The largest uncertainties are about 2% and 3% for Rn and Tn respectively. The minimum detection limits are around 2.0 Bq/m3 for both gases. If only Rn concentration is required, the measurement time can be significantly shortened.
Journal title :
Applied Radiation and Isotopes
Journal title :
Applied Radiation and Isotopes