Title of article :
Dosimetry for optimisation of patient protection in computed tomography
Author/Authors :
K. A. Jessen، نويسنده , , P. C. Shrimpton، نويسنده , , J. Geleijns، نويسنده , , W. Panzer، نويسنده , , G. Tosi، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1999
Pages :
8
From page :
165
To page :
172
Abstract :
The complex conditions of irradiation in computed tomography (CT), involving highly-collimated X-ray beams, necessitate the use of specially-defined dose descriptors such as the computed tomography dose index (CTDI). When used in a weighted form (CTDIW), this concept can describe the absorbed dose from a single slice in standard head and body phantoms. The model can easily be extended to characterise patient exposure for a complete examination by means of the reference dose quantity dose-length product (DLP). Effective dose can also be estimated from DLP, when required.
Keywords :
radiation dose , dosimetry , computer assisted tomography
Journal title :
Applied Radiation and Isotopes
Serial Year :
1999
Journal title :
Applied Radiation and Isotopes
Record number :
540296
Link To Document :
بازگشت