Title of article :
Automated synthesis of [11C]choline, a positron-emitting tracer for tumor imaging
Author/Authors :
Toshihiko Hara، نويسنده , , Mitsuaki Yuasa، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1999
Pages :
3
From page :
531
To page :
533
Abstract :
(β-Hydroxyethyl)tri([11C]methyl)ammonium ([11C]choline) is a tracer very effective in imaging various human tumors using positron emission tomography (PET). We have constructed a computer-controlled [11C]choline synthetic apparatus which carries out the whole process of synthesis and product purification automatically. The setup is simple and the process quick. In 20 min, 11 GBq of [11C]choline (chloride) is obtainable from 26 GBq of [11C]CO2. The final product is a sterile and pyrogen-free [11C]choline “injection”.
Journal title :
Applied Radiation and Isotopes
Serial Year :
1999
Journal title :
Applied Radiation and Isotopes
Record number :
540339
Link To Document :
بازگشت