Title of article
Evaluation of peak-to-total ratio for germanium detectors
Author/Authors
Tamaki Watanabe، نويسنده , , Yoshihiro Oi، نويسنده , , Mitsumasa Taki، نويسنده , , Katsuya Kawasaki، نويسنده , , Makoto Yoshida، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1999
Pages
5
From page
1057
To page
1061
Abstract
An equation for the peak-to-total ratio, P, of a germanium detector is proposed in the following form. P=var epsilonp/var epsilont+a(V/A)+b(V/A)2/2 where var epsilonp/var epsilont is the ratio of the photoelectric to the total detection efficiency for a germanium detector, V/A is the volume-to-surface ratio of the detector, and the parameters of a and b are experimentally given as a function of incident photon energy. The first term in the right side of the equation expresses the fractional single photoelectric-to-total interaction between photons and the detector. The second and third terms correspond to the multiple interactions. The peak-to-total ratios for the germanium detector with a sensitive volume up to 200 cm3 were calculated by the equation. They coincide with the experimental values to within an uncertainty of several percent in the energy region 0.3–3 MeV.
Keywords
germanium , radiation detector , cobalt 60 , cesium 137 , sodium 24 , chromium 51
Journal title
Applied Radiation and Isotopes
Serial Year
1999
Journal title
Applied Radiation and Isotopes
Record number
540406
Link To Document