Abstract
This article outlines a simple experimental procedure to obtain the minimum biased detection efficiency calibration for gamma waste assay systems using the so called shell-source method in cylindrical waste drums. The conventional use of the shell-source method to simulate a homogeneous activity distribution in a cylindrical waste drum results in a systematic underestimation of the detection efficiency. The bias may be more then 30% at gamma energies of 59 keV and a waste density of 2 g/cm3. Experimental evidence of this bias is given for the low level waste assay system, Q2, and a simple measurement procedure to eliminate this bias is presented. The proposed method does not rely on data from computer simulations and is applicable to all types of waste matrices.
Keywords
Radioactive waste disposal , radioisotope