Title of article :
Mixing activity-calibrated and uncalibrated gamma-ray sources in efficiency calibration
Author/Authors :
R. Tramontano، نويسنده , , V. R. Vanin، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1999
Abstract :
A method to calibrate the absolute efficiency of HPGe detectors in a large energy range using both activity-calibrated and uncalibrated multienergy gamma-ray radioactive sources is presented. The adopted efficiency calibration function var epsilon(E) has the simple form ln var epsilon(E)=a0+a1 ln(E/EB)+a2 ln2(E/EB). Different parameters a0 are simultaneously fitted to the activity-calibrated and uncalibrated sources experimental efficiency data by the least-squares method. As an application example, the calibration of a HPGe detector in the 500 to 3500 keV energy range using a 207Bi activity calibrated source and one uncalibrated 56Co source is described.
Keywords :
gamma radiation
Journal title :
Applied Radiation and Isotopes
Journal title :
Applied Radiation and Isotopes