A hybrid radioassay technique for multiple beta-emitter mixtures using liquid and NaI(Tl) scintillation spectrometers
Author/Authors
Makoto Takiue، نويسنده , , Takashi Natake، نويسنده , , Haruo Fujii، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1999
Pages
6
From page
429
To page
434
Abstract
Multiple pure-beta and beta-gamma emitter mixtures can be reliably radioassayed using a combination of a liquid and a NaI(Tl) scintillation spectrometer. The method of least squares has been applied to the radioassay of relatively low-activity samples. This system removes the difficulty of analysing beta-emitter mixtures with very similar liquid scintillation pulse height distributions, such as 3H, 51Cr and 125I. All that is required for the technique is a quenched standard sample set of the nuclides to be analyzed and calibration standard samples for the gamma-emitters.
Keywords
Gamma-ray spectrometer , Liquid scintillation spectrometry , Assay beta emitters