Title of article :
Cryogenic detectors below 100 mK for X-ray measurements in metrology
Author/Authors :
C. Bobin، نويسنده , , E. Leblanc، نويسنده , , J. Bouchard، نويسنده , , N. Coron، نويسنده , , P. Cassette، نويسنده , , J. Leblanc، نويسنده , , P. de Marcillac، نويسنده , , J. Plagnard، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2000
Pages :
4
From page :
403
To page :
406
Abstract :
Due to the intrinsic performances of cryogenic detectors such as energy resolution, LPRI has decided to use these devices to improve the quality of the radioactive measurements usually obtained with classical semiconductor detectors. A bolometer with a 10 μg tin absorber has been developed at IAS (Institut d’Astrophysique Spatiale) and has been tested in the cryogenic installation of LPRI; an energy resolution (full width half maximum, FWHM) of 39 eV has been obtained on the Kα line of Mn. Besides these good spectrometry results, an absolute activity measurement using bolometers is proposed by adapting an absorber geometry for 4 π counting.
Journal title :
Applied Radiation and Isotopes
Serial Year :
2000
Journal title :
Applied Radiation and Isotopes
Record number :
540565
Link To Document :
بازگشت