Title of article
Accuracy considerations in EPR dosimetry
Author/Authors
Vitaly Nagy، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2000
Pages
12
From page
1039
To page
1050
Abstract
A brief overview of some of the potential sources of inaccuracies that are common to all kinds of electron paramagnetic resonance (EPR) dosimetry is given. The instrumental sources discussed are inaccuracies of receiver gain calibration, instability of the spectrometer sensitivity, and neglect of inhomogeneity of the fields in the resonators. The discussed sources of inaccuracies related to data processing are violations of basic assumptions of the used version of the least-squares technique, incorrect model selection for fitting the data, and nonoptimal design of the calibration experiment. Methods for detecting and correcting such inaccuracies are briefly discussed and references to the original literature are given.
Journal title
Applied Radiation and Isotopes
Serial Year
2000
Journal title
Applied Radiation and Isotopes
Record number
540653
Link To Document