Title of article :
Transferability of ASTM/NIST alanine–polyethylene recipe at ISS
Author/Authors :
C. De Angelis، نويسنده , , P. Fattibene، نويسنده , , S. Onori، نويسنده , , E. Petetti، نويسنده , , A. Bartolotta، نويسنده , , A. Sansone Santamaria، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2000
Pages :
5
From page :
1197
To page :
1201
Abstract :
Alanine–polyethylene solid state dosimeters were prepared at Istituto Superiore di Sanità (ISS) following the recipe proposed by National Institute of Standards and Technology (NIST) with the goal of testing its transferability. Dosimeters were prepared using 95% alanine and 5% polyethylene, by weight. They are rugged and of increased sensitivity, repeatability and reproducibility as respect to the ISS alanine-paraffin pellets. Reproducibility of about 1% was obtained at 10 Gy and at 3 Gy if one single pellet or a stack of five dosimeters were used, respectively.
Journal title :
Applied Radiation and Isotopes
Serial Year :
2000
Journal title :
Applied Radiation and Isotopes
Record number :
540679
Link To Document :