Title of article :
Monte Carlo calculation of the efficiency calibration curve and coincidence-summing corrections in low-level gamma-ray spectrometry using well-type HPGe detectors
Author/Authors :
J. -M. Laborie، نويسنده , , G. Le Petit، نويسنده , , D. Abt، نويسنده , , M. Girard، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2000
Abstract :
Well-type high-purity germanium (HPGe) detectors are well suited to the analysis of small amounts of environmental samples, as they can combine both low background and high detection efficiency. A low-background well-type detector is installed in the Modane underground Laboratory. In the well geometry, coincidence-summing effects are high and make the construction of the full energy peak efficiency curve a difficult task with an usual calibration standard, especially in the high energy range. Using the GEANT code and taking into account a detailed description of the detector and the source, efficiency curves have been modelled for several filling heights of the vial. With a special routine taking into account the decay schemes of the radionuclides, corrections for coincidence-summing effects that occur when measuring samples containing 238U, 232Th or 134Cs have been computed. The results are found to be in good agreement with the experimental data. It is shown that triple coincidences effect on counting losses accounts for 7–15% of pair coincidences effect in the case of 604 and 796 keV lines of 134Cs.
Keywords :
Well-type HPGe detectors , Monte Carlo simulation , Efficiency calibration , Gamma-ray spectrometry , True coincidence-summing
Journal title :
Applied Radiation and Isotopes
Journal title :
Applied Radiation and Isotopes