    Energy linearity of high-purity germanium detectors in the region of the Ge K-absorption edge: experimental results

  • Author/Authors

    J. M. F. Dos Santos، نويسنده , , C. M. B. Monteiro، نويسنده , , R. E. Morgado، نويسنده , , C. A. N. Conde، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2000
  • Pages
    5
  • From page
    739
  • To page
    743
  • Abstract
    Non-linearities in the energy response of gas detectors in the regions of the absorption edges of the detection medium are well documented. Monte Carlo calculations show that the non-linearity results from differences in efficiencies for converting absorbed radiation into ionisation for different atomic sub-shells. Energy non-linearity in germanium-based solid-state detectors in the region of the germanium 11.104-keV K-edge is not well documented, although a 1% non-linearity has been previously reported in a Ge(Li) detector. This relatively high value is of practical concern since high-purity germanium (HPGe) is often the detector of choice for α- and X-ray spectrometry down to a few keV. In this paper, we present the experimental results for the energy response of a HPGe detector in the 8–15-keV energy region of the germanium K-edge. Within the accuracy of our measurements, we conclude that there is no measurable non-linearity effect in germanium at the K-edge.
  • Keywords
    Energy linearity , HPGe , K-edge